6. ECOLOG-NET LP4 / WP4
6.1 Operating datalogger type ...P4
Display „On"
Error 5; module does not respond during data read-out
- The datalogger is not in measuring mode
- Press F1/F3 to exit the menu
6.2 Functions of collective alarm contact; datalogger type ...P4
Datalogger type ECOLOG-NET ...P4, has 4 possibilities for alarm signaling:
Relay contact S
Semiconductor contact Alarm 1
Semiconductor contact Alarm 2
Internal buzzer
D-EL-6003Bk
Manual
ECOLOG-NET
4
only appear when the datalogger is in
power-save mode and is logging data.
Press the menu key to activate the display for a
few seconds.
Start
Only appears when the datalogger is in start/stop
mode and is waiting for the start time.
Stop
Status codes
Resets alarm display and contacts, causes a reset
timestamp in the alarm protocol
Appears if an alarm is active or the "self-sustaining"
function becam selected and an alarm has been logged.
(
Chapter 5)
No function (exit menu)
Time stamp indicated as D2
(
Chapter 2.12)
LCD display test
OFF
OFF
ON
Alarm output test
This is a potential-free switch-over contact. The contact switches when an
alarm is triggered or when the external power supply fails.
Active during an alarm only
This is a semiconductor switch - break contact to datalogger GND
Active during an alarm only
This is a semiconductor switch - make contact to datalogger GND
Acoustic alarm signal (
Chapter 2.10)
Exit menu