Electrical connection
5.1.1

Mechanical connection of the connector cables

The junction box includes a set of cable sleeves, intended for bushing the
transducer connection cable, the voltage supply and the Md/n outputs. The
side hole for the external control signal is closed with a screw for delivery. If
need be, use a suitable PG7 screwed joint with an anti‐buckling socket.
T20WN, T21WN, T22
Anti‐buckling socket,
internal Ø 7 mm
Power
Anti‐buckling
socket, internal
Ø 5 mm
Fig. 5.3
Cable bushing "T20WN, T21WN, T22" and "POWER" at the junction box
12
Clamping ring
Cable sleeve
Rubber seal
A02335_03_Y00_00 HBM: public
Cable shield
looped round
clamping ring
Wall of the housing