Preparing the Scale for Use
Selecting Installation Location
The correct location is crucial to the accuracy of the weigh-
ing results.
•
Select a stable, vibration-free and, if possible, a hori-
zontal location for the weighing platform.
•
The ground must be able to safely bear the weight of
the fully loaded weighing platform.
Observe the following environmental conditions:
•
– No direct sunlight
– No strong drafts
– No excessive temperature fluctuations
Installing the Scale
•
Place the scale at the installation location.
•
Check whether the rocker-feet are installed in the 4
load cells. Do not apply torque to the load cells.
•
Route the cable to the weighing terminal in such a
way that it is protected against damage.
•
For further steps refer to the weighing terminal user
manual.
1
3
1. Platform
2. Side board
3. Junction box
4. Load cell
5. Washer
Fig. 3-1 Major Components
4
5
6
7
6. Bolt
8
7. Spring
8. Load cell rocker-foot
9. Expanding bolt
10. Foot plate
METTLER TOLEDO
9
10
9
User Manual BUA221/BUA226