Endress+Hauser micropilot FMR 230 V Sicherheitshinweise Seite 8

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 3
en XA 020F-A
Electronics compartment
EEx ia
HART electronics
Tank
Hazardous
area
Zone 0
Ue ≤ 30 V
Power supply
Um ≤ 250 VAC
Housing
T12
Version
FMR 230V-........C
FMR 230V-........D
FMR 230V-........F
Safety notes for installation in hazardous areas:
1) Install according to the manufacturer's instructions and other valid standards and guidelines.
2) The Micropilot is to be connected to the local equipotential bonding system.
The input circuit is galvanically connected to the housing.
3) The dependency of the permissible ambient temperature of the electronic housing upon the application and temperature
class is given in Tables 1 and 2.
4) If the housing is realigned, check that the locking screw is screwed tight again, see operating manual.
5) A label must be attached to the terminal compartment:
" Do not open in explosive atmospheres when the power is on ".
2
Explosion hazardous area
Connection compartment (EEx e)
5
Electronics compartment
EEx ia
HART electronics
Tank
Hazardous
area
Zone 1
Plant grounding system (here German practice)
Connection
compartment (EEx e)
–40 °C ≤ Tamb ≤ 75 °C
–40 °C ≤ Tamb ≤ 200 °C
–40 °C ≤ Tamb ≤ 200 °C
–40 °C ≤ Tamb ≤ 350 °C
5)
See note
Optionally with or without operating and display module VU 330
For use see table
For use see table
For use see table
Micropilot FMR 230V
Non-hazardous area
Power supply
Endress+Hauser