Problem
Possible cause
Pulse oximeter
Circulation problems in the
shows interrupted
finger being measured.
measurements or
The finger being measured is
severe jumps in the
too large or too small.
measurement.
The finger, hand or body are
moving.
Arrhythmia
Solution
Refer to warnings and safety notes.
The fingertip must be the following size:
between 10 and 20 mm wide
between 5 and 15 mm thick
Keep the finger, hand and body still during
measurement.
Seek medical attention.
GB/IE