Herunterladen Inhalt Inhalt Diese Seite drucken

Low Battery; Keypad Lock - TensCare Sure+ Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

sudden feeling of a surge, as each
programme gives a different sensation.
TREATMENT TIMER
The button marked T can be
T
used to set the session
duration. When you switch the unit on,
it is automatically set at 20 min or
Continuous,
(displayed
depending on the programme chosen.
To set a different time, set the strength
to zero and press T. The min display
will flash.
You can set session times of C
(Continuous) or 10, 20, 30, 45, 60 or 90
minutes with the buttons ▲ and ▼.
Press T again to save your selection.
The LCD shows the session duration
next to the clock symbol. The unit
automatically counts down the minutes
set and switches off when it reaches 0.

LOW BATTERY

An empty battery symbol will
show when you need to
change the batteries. The unit
will shut down about 2 minutes after
this.

KEYPAD LOCK

If you do not press any keys
for 30 seconds, the keypad
will lock. This is to avoid
accidental changes in setting.
To unlock, press and hold the strength
down control (button ▼).
LEADS ALARM
The Sure+ monitors the connection and
the contact between you and the probe,
or the pads. This is to prevent sudden
changes if a broken connection is re-
made. If either of these goes outside of
a standard range while the strength is
above 20, the unit will flash LEADS and
return the strength to 0.
Check the lead and if necessary,
lubricate the probe with a water based
lubricant such as TensCare Go Gel
(see K-GO). Please see section 20 for
as
C)
more troubleshooting tips.
MEMORY
The Sure+ has a Memory with three
functions:
1) Programme Retention. When you
switch the unit on, it will automatically
start in the programme which was being
used when it was switched off.
2) Usage. Press T and ▼ together and
hold down for 3 to 5 seconds. The
display will show the number of times
the unit has been used and the duration
of use in hours.
Press the same buttons again to return
to normal controls.
3) Memory Reset. To reset memory to
zero, hold down the T and OFF buttons
together for 3 to 5 seconds.
21
Number of
uses
Time used
in hours

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis