Each time you change the programme,
the strength level reverts back to zero.
This is a safety feature to alleviate any
sudden feeling of a surge, as each
programme gives a different sensation.
TREATMENT TIMER
The button illustrated can be
used
to
duration. When you switch the
unit on, it is automatically set at 20 min
or
Continuous,
(displayed
depending on the programme chosen.
To set a different time, set the strength
to zero and press the TREATMENT
TIMER button. The min display will
flash.
You can set session times of C
(Continuous) or 10, 20, 30, 45, 60 or 90
minutes with the buttons ▲ and ▼.
Press the button again to save your
selection.
The LCD shows the session duration
next to the clock symbol. The unit
automatically counts down the minutes
set and switches off when it reaches 0.

KEYPAD LOCK

The Alivia keypad can be
locked
by
KEYPAD LOCK button. If you do not
press any keys for 30 seconds, the
keypad will auto-lock. This is to avoid
accidental changes in setting. To
unlock, press keypad lock button.
LEADS ALARM
The Alivia monitors the connection and
the contact between you and the probe,
or the pads. This is to prevent sudden
changes if a broken connection is re-
set
the
session
as
C)
pressing
the
made. If either of these goes outside of
a standard range while the strength is
above 20, the unit will flash LEADS,
bleep three times, and return the
strength to 0.
Check the lead and if necessary,
lubricate the probe with a water based
lubricant such as TensCare Go Gel
(see K-GO). Please see section 20 for
more troubleshooting tips.

LOW BATTERY

An empty battery symbol will
show when you need to
change the batteries. The unit
will shut down about 2 minutes after
this.
MEMORY
The Alivia has a Memory with three
functions:
1) Programme Retention. When you
switch the unit on, it will automatically
start in the programme which was being
used when it was switched off.
2) Usage. Press the MEMORY
button. The display will show
the number of times the unit
has been used and the duration of use
in hours.
Press the button again to return to
normal controls.
3) Memory Reset. To reset memory to
zero, hold down the MEMORY and
ON/OFF buttons together for 3 to 5
seconds.