PicoScope 6000E Series oscilloscopes
2.2 Maximum input/output ratings
Observe all terminal ratings and warnings marked on the product. The marked voltage is the maximum that may be
applied across that terminal without risk of personal injury or damage to the instrument.
The table below and markings on the product indicate the full-scale measurement range and overvoltage protection
range for each oscilloscope model. The full-scale measurement ranges are the maximum voltages that can be
accurately measured by the instrument. The overvoltage protection ranges are the maximum voltages that will not
damage the instrument.
To prevent electric shock, do not attempt to connect to or measure voltages outside of the marked terminal rating.
All models
measurement range
1 MΩ input
50 Ω input
Signals exceeding the voltage limits in the table below are defined as "hazardous live" by EN 61010.
To prevent injury or death, do not directly connect the oscilloscope to hazardous live voltages.
To prevent electric shock, take all necessary safety precautions when working on equipment where hazardous live
voltages may be present.
The following accessories can safely connect to and allow measurement of hazardous live voltages, up to either the
instrument's full-scale measurement voltage multiplied by the applicable attenuation ratio or the accessory's marked
maximum operating voltage, whichever is the lower:
P2036 and P2056 10:1 passive high-impedance oscilloscope probes
All Pico Technology-supplied high-voltage active probes
Accessories suited for connection to hazardous live voltages are clearly marked with their maximum allowable
voltage and, where applicable, their CAT rating.
To prevent electric shock, do not exceed the voltage rating marked on any accessory. If an accessory is not
marked with a voltage rating on either the connector, cable or body, or if a protective finger guard is removed, then
do not exceed the EN 61010 "hazardous live" limits above. When connecting one or multiple accessories and an
instrument channel together, the lowest voltage rating in the chain applies to that channel.
To prevent injury or death, do not directly connect the oscilloscope to the mains (line power). To measure mains
voltages, use only a differential isolating probe that is specifically CAT rated for mains or high energy use, such
as the TA041 listed on the Pico website. Such measurements must only be made by competent persons and
appropriate safety training may be required.
4
Full-scale
Input channels
±20 V
±100 V
5.5 V RMS,
±5 V
±10 V peak max.
Signal voltage limits of EN 61010-1: 2010+A1:2019
±60 V DC
Copyright © 2020–2021 Pico Technology Ltd. All rights reserved.
WARNING
Overvoltage protection (DC + AC peak)
Aux trigger
±20 V
WARNING
30 V AC RMS
WARNING
WARNING
10 MHz reference
Signal generator
clock input
±5 V
±42.4 V pk max.
User's Guide
±20 V
DO338-5