Herunterladen Inhalt Inhalt Diese Seite drucken

Troubleshooting Table - SCHÖLLY FIBEROPTIC FLEXILUX 200 Gebrauchsanweisung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

6.6

Troubleshooting Table

Issue
No operation
Light doesn't switch on
The light guide cannot be
coupled
Not enough light
Fan doesn't work
Brightness significantly too
low
Light source cannot be con-
trolled
Instructions For Use | 13 - Mar - 2020 | Version: H
Possible causes
No power supply
Malfunction in power supply unit
Device is overheated
Shutter closed/light off
Adapter on light guide connector
missing
Light guide not connected cor-
rectly
Light guide defective
Adapter not screwed on correctly
Malfunction in fan
Light guide not completely insert-
ed
Interface cable not correctly
plugged in or defective
Table 6-1: Troubleshooting table.
6 Maintenance and Repair
6.6 Troubleshooting Table
Fixes
Check the mains connection and
connect the power if necessary
Check the fuse and replace if neces-
sary
Send the light source in for repair
Allow device to cool down
Open the shutter/switch on light
Screw on adapter
Check that the light guide is posi-
tioned properly; correct if necessary
Replace the light guide
Screw on the adapter
Send the light source in for repair
Use the correct light guide connec-
tion
Check connection or replace cable
Page 70 of 76