Combination with Medical Electrical Equipment

2.2.5

Display of Faults

Instructions For Use | 13 - Mar - 2020 | Version: H
The device can be combined with components from other manufacturers provided
that all components are compliant with the medical electrical equipment safety re-
quirements according to IEC 60601-1.
It is the operator's responsibility to check and make sure that the system is and
remains fully operational.
When using devices from different manufacturers and when operating an endo-
scope and/or endoscopic accessories with electromedical devices, it must be en-
sured that the applied part is properly insulated (in accordance with the require-
ments for type BF/CF applied parts defined in IEC 60601-1).
Contact our Service Department if you have any questions regarding compatibility.
No light guide or light guide attached incorrectly
If the light guide is not attached or is not attached correctly, you will see the fol-
lowing on the display:
Figure 2-1: Symbols displayed if light guide is not attached.
If this occurs, you should attach the light guide, attach it correctly, or switch the
device off.
Operating error in LED module
In case of an operating error in the LED module, the display shows the following:
Figure 2-2: Message displayed on operating error in LED module.
If this occurs, you should replace the light source unit as quickly as possible.