2 General Safety Information
2.2 Product Safety
Failure of housing fan
If the housing fan fails, the following message is displayed:
Figure 2-3: Message displayed when housing fan fails.
When housing fan fails, the temperature inside the device climbs in relation to the
ambient temperature. If the ambient temperature is high, the device may overheat
after 30 to 45 minutes.
If this occurs, you should replace the light source unit as quickly as possible.
Overheating
The device features an internal temperature sensor.
In very rare cases, for instance if the fan in the device is obstructed, the internal
temperature of the device could rise above 60 °C.
If the temperature rises above 60 °C, you will see the following warning on the
display:
Figure 2-4: Warning displayed when temperature exceeds 60 °C.
In order to protect the LED, the device will, in the event of overheating, limit the
maximum light output to 70 %. The device will not switch off, but is not guaran-
teed to remain fully functional.
If this occurs, you should replace the light source unit as quickly as possible.
Instructions For Use | 13 - Mar - 2020 | Version: H