Note:
When using the UPPER Z and LOWER Z keys to
go to a threshold, the keys have to be pressed
until the final position is reached.
There are two step sizes: Fine and Coarse.
In the Fine mode, the value is adjusted to the ob-
jective in use. The values are pre-defined. The
settings can be changed with the Software
Leica Application Suite, module: Fine Tuning.
In the Coarse mode, the value is the same for all
objectives. Coarse represents the maximum ve-
locity.
Note:
When you assign a certain step size to an objec-
tive, this does not only affect the Z-drive but also
the stage-step size that is assigned to this ob-
jective when selecting the Precise mode.
Switch between Fine and Coarse via:
Leica SmartTouch – navigation item
Variable function keys at the stand and at
the SmartMove
Leica STP6000
Software Leica Application Suite, module:
Acquire

8.4 Tubes

Note:
Close any unused tube openings, as otherwise
stray light can interfere with observation.
Note:
Make sure that the connector cable is plugged
in on the MBDT25+ motorized tube (67.1).
Adjusting the Viewing Distance
• Adjust the viewing distance of the eye-
pieces so that a congruent total image is seen
(Fig. 67).
Adjusting the Viewing Angle
• For the AET22 and EDT22 ergotubes, the view-
ing angle can be adjusted by tilting the bin-
ocular viewer in the range of 5° - 32° (Fig. 68)
Fig. 67
Tube setting
↔ Personal eyebase settings
1
Motorized tube connection
1
8. Operation