Caution!
Through activating to the grounding connec-
tion (earth screw on the back of the Leica
CTR5500 and CTR6000 Electronics Box)
ancillary equipment with its own and/or
extra power supply may be brought to the
same ground wire potential. For connections
without a ground connector, Leica Service
must be consulted.
Caution!
Never use any fuses as replacements other
than those of the types and the current rat-
ings listed here. Using patched fuses or
bridging the fuse holder is not permitted.
Caution!
The microscope's electrical accessory com-
ponents are not protected against water.
Water can cause electric shock.
Caution!
Protect the microscope from excessive tem-
perature fluctuations. Such fluctuations can
lead to the accumulation of condensation,
which can damage the electrical and optical
components.
Ambient temperature: 15-35°C.
Caution!
Before exchanging the fuses or lamps, be
absolutely certain to switch-off the main
power switch and remove the power cable.
Caution!
Touch the touchscreen using your finger
only. Never use a pen or other hard, sharp or
pointed objects.

3.3 Disposal

After the end of the product's life, please
contact Leica Service or Leica Sales on how to
dispose of it.
Please
observe
the
ordinances which, for example, implement and
ensure compliance with EU directive WEEE.
n.b.:
Like
all
electronic
microscope, its components and expend-
ables may not be disposed of as general
household waste!
3. Safety Notes
national
laws
and
instruments,
the
11