5.28 Opening the Door in Case of Power Failure

If there is a power failure, ensure the machine is isolated from the main power supply.
Then remove the kickplate of the HYDRIM. Remove the screw with a screwdriver
and lift the kickplate up and toward you. Locate the ring and pull it to open the door.
Exercise caution! There may be fluid remaining in the unit and the instruments may be
hot. Instruments that have not completed the cycle should not be used and should be
reprocessed. When replacing the kickplate, ensure that the grounding wire is attached.
Kickplate
Cycle count
This allows the user to see how many cycles have been run on
the HYDRIM. Touch cycle count and the number of cycles
(completed and aborted) will be displayed. This cannot be reset.
Touch the back arrow to return to the previous menu.
Instructions
Users can access a database of pictorial instructions on how to
perform routine maintenance such as installing the cleaning solution.
Screw
Ring
28
Grounding wire