Technical data
Intervals before readiness for operation
1st measurement
after a previous measurement with:
0.25 mg/L
0.50 mg/L
1.00 mg/L
> 1.5 mg/L
Safety inspection
Calibration
Functional test/
Quality assurance
Dimensions
Weight
36
approx. 6 s
20 s
40 s
80 s
120 s
every 6 months
every 6 months
Recommended daily. Quality assurance
must be performed in accordance with
§4a MPBetreibV (German act on the
installation, operation and use of
medical devices) and "RiliBÄK zur
Qualitätssicherung laboratoriumsmedi-
zinischer Untersuchungen" (Guideline
of the German Medical Association on
quality assurance for medical laboratory
examinations), part B1, 2.1.1.
80 mm x 140 mm x 35 mm (W x H x D)
approx. 195 g
Electrical supply
Batteries
Rechargeable batteries
Interface
Classification:
per EC Directive 93/42/EWG,
Annex IX
UMDNS-Code
DIMDI registration number
2 pieces 1.5 V Type Alkaline
(Mignon, LR6, AA)
approx. 1500 measurements, or
2 pieces 1.2 V NiMH
(Mignon, LR6, AA)
RS 232,
for data transfer to PC
Measuring instrument:
Class I with a measuring function
Mouthpieces:
Class I
17-475
DE/CA01/102/004
Dräger Alcotest 6810 med