Handling
Handle with care especially when fitted with the extra thin and sharp spring
contact tip to avoid any injury. Note that the probe cable is a sensitive part of the
probe. Do not damage through excessive bending or pulling. Avoid mechanical
shock to this product in general to guarantee accurate performance and
protection.
Use ground lead only for connections to earth ground.
The accessories provided with the probe have been safety tested. Do not use any
other accessories than those "originally" provided.

Maintenance

● Cleaning
To clean the exterior of the probe use a soft cloth moistened with either distillated water or
isopropyl alcohol. Before use allow the probe to dry completely.
● Changing the Probe Tip
To change the probe tip use pliers to grip and pull it carefully straight out of its contact socket,
along the axis of the probe. Do not grip the white plastic insulator or the housing with pliers, be-
cause the tip could be squeezed and cannot be removed and respectively the probe could be da-
maged. If the probe tip is removed, the new tip can be inserted with pliers into the contact socket,
along the axis of the probe. In order to insert the probe tip completely into the housing, press the
probe tip against a hard surface carefully.
Most oscilloscopes provide a build-in function generator to verify the passive probe is compen-
sated correctly. After changing the probe tip always use the build-in function generator or other
stand-alone instrument to make sure the probe is safe to operate.
Use pliers to grip and pull the probe tip carefully
out of its contact socket.
Do not grip the white plastic insulator or the probe
housing with pliers.
PML SERIES
EN
33