Caretalk DSO 364 Gebrauchsanweisung Seite 55

Stirn-/ohrthermometer
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Rated max-
Separation distance according to
imum out-
frequency of transmitter m
put power
0.15 MHz to
of transmit-
80 MHz
ter W
0.01
0.12
0.1
0.38
1
1.2
10
3.8
100
12
For transmitters rated at a maximum output pow-
er not listed above, the recommended separation
distance d in meters (m) can be estimated using
the equation applicable to the frequency of the
transmitter, where P is the maximum output pow-
er rating of the transmitter in watts (W) according
to the transmitter manufacturer.
NOTICE 1 In the case of 80 MHz and 800 MHz, the
higher frequency range shall apply.
NOTICE 2 These guidelines may not apply in all
situations. Electromagnetic propagation is affected
by absorption and reflection from structures, ob-
jects and people.
Electromagnetic immunity (EMI)
80 MHz to
800 MHz
0.12
0.38
1.2
3.8
12
80 MHz to
2.7 GHz
0.23
0.73
2.3
7.3
23
55