Explanation – electromagnetic immunity
The device is intended for use in the electromag-
netic environment specified below. The customer
or the user of the device should ensure that it is
used in such an environment.
Immunity
IEC 60601
tests
test level
Electro-
± 8 kV
static
contact
discharge
± 2 kV,
IEC
± 4 kV,
61000-4-2
± 8 kV,
± 15 kV air
Fast
± 2 kV for
transient
power
electrical
supply
distur-
cables
bances/
± 1 kV for
bursts
input/out-
IEC
put cables
61000-4-4
Electromagnetic immunity (EMI)
Table 2
Com-
pliance
level
± 8 kV
contact
± 2 kV,
± 4 kV, ±
8 kV, ±
15 kV air
Not ap-
plicable
Electromagnet-
ic environment
– instructions
Floors should
be wood, con-
crete or ceramic
tile. If floors
are covered
with synthetic
material, the
relative humid-
ity should be at
least 30 %.
Outlet power
quality should
be that of a typ-
ical commercial
or hospital
environment.
49