Troubleshooting

Troubleshooting
Fault
The bat-
teries have
been insert-
ed, but the
LCD display
is blank.
1
44
Cause
There may
be debris in
the battery
compartment.
The batteries
are empty or
were inserted
with incorrect
polarity.
The battery
contacts are not
aligned with
the terminals.
There is a
fault in the
electronics.
The tempera-
ture measured
is less than
32.0 °C/89.6 °F
or higher than
42.9 °C/109.3 °F.
The thermom-
eter is defective
(hardware
error).
Solution
Remove the
debris from
the battery
compartment.
Replace the
batteries with
full batteries.
Make sure
the polarity is
correct.
Check the bat-
tery contacts.
Remove the
batteries and
put them back
in after approx.
3 seconds.
Observe the
notices in the
user manual.
Contact the
manufacturer/
vendor.