Herunterladen Inhalt Inhalt Diese Seite drucken

Rear Panel - Rohde & Schwarz Hameg HM8123 Handbuch

3 ghz programmable counter
Vorschau ausblenden Andere Handbücher für Hameg HM8123:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

nd
2
Setting the trigger level with the numerical keys
the entered value will be accepted with the unit millivolts
(mV).
ARMED / MHz (pushbutton)
Button with two functions:
st
1
Pressing this button will select the ARMED mode.
nd
2
Setting the gate time with the numerical keys
entered value will be accepted with the unit milliseconds
(ms).
OFFSET / Hz/ns (pushbutton)
Pressing this button will activate the OFFSET function.
GATED / kHz/us (pushbutton)
Pressing this button will activate the GATED mode.
Function pushbuttons
to
These buttons have two functions:
st
1
The measurements function are called by operating these
buttons. The corresponding button will light up.
nd
2
Setting the trigger level or the gate time, the desired
value can be entered with this buttons and the unit buttons
(mV
, V
and ms
, s
). Also refer to chap. Gate time
and Triggering. None of these button is lit.
FREQ A
Frequency channel A
FREQ B
Frequency channel B
FREQ C
Frequency measurement channel C
PER A
Period channel A
Width A
Pulse width channel A
A : B
Frequency ratio of channels A : B
Duty A
Duty cycle channel A
TI A B
Time interval A B
TI avg A B Time interval A B averaged
,
Phase A B Phase difference A B
RPM A
TOTAL A
TRIG/ARM INPUT (BNC connector)
External gate control input
, the

Rear Panel

Interface
RS-232 interface (Options: IEEE-488, USB)
A (BNC connector)
Trigger signal channel A output (e.g. for display on a scope).
Signal level 0 to + 5 V (TTL level)
B (BNC connector)
Trigger signal channel B output (e.g. for display on a scope).
Signal level 0 to + 5 V (TTL level)
GATE (BNC connector)
Gate view output. This output will be high as long as the gate
is open during a measurement.
10 MHz Ref. (BNC connector)
External reference input (10 MHz)
RESET (BNC connector)
External reset signal input (TTL level). The function is iden-
tical to that of the RESET pushbutton
Mains input connector
C o n t r o l s a n d d i s p l a y
Rpm measurement channel A
Event counting channel A
.
Subject to change without notice
23

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Hameg hm8123g

Inhaltsverzeichnis