Valor™ 2000W Series

3.4 Initial Calibration

When the scale is first installed it should be calibrated to ensure accurate results.
Before performing the calibration, be sure to have the appropriate calibration
weights as listed in table 3-3. Press and hold Cal for 5 seconds, the display will
show [Cal]. When the button is released, the display will show [SpaN]. Press
On/Zero to begin the span calibration. [--C--] blinks while zero reading is stored.
Next, the display shows the calibration weight value.
Place the specified calibration mass on the pan and press On/Zero. [--C--] blinks
while the reading is stored. The scale returns to the previous application mode and
is ready for use.
Err 3.0 Cal
The message [
] will be displayed if these calibration steps are not
followed or if the wrong weight was used.
The calibration process can be aborted by turning the scale off.
TABLE 3-3
Required Span Calibration Mass
(sold separately)
Max
Mass
6000 g
5 kg
15000 g
10 kg