Valor™ 2000W Series
3.2 Turning Scale On/Off
To turn the scale on, press the On/Zero Off button. The scale performs a display
test, momentarily displays the software version, and then enters the active weighing
mode.
To turn the scale off, press and hold the On/Zero Off button until OFF is displayed.

3.3 Calibration Menu

To enter the calibration menu, extend press the Cal button for 5 seconds, [SpaN] will
then be displayed. Press Zero key to perform span calibration or press Cal key to
proceed to GEO settings. Press Zero key to enter GEO settings and then press Cal
key to increase the value. Press Zero key to confirm, [end] will be displayed. Press
Zero key to exit the menu and return to weighing mode.
Cal
Span
GEO
End
Figure 3-4. Menu structure
Span [SpaN] (yes, no)
Initiates a span calibration procedure (zero and span).
GEO [GEO]
Geographical Adjustment Factor (GEO) is used to adjust the calibration based on
the current location. Settings from 0 to 31 are available with 12 being the default.
Refer to table 4-2 to determine the GEO factor that corresponds to your location.
Figure 3-3. Valor 2000 – V22PW Rear display.
EN-5