Plug
CO
2 absorber
testo 400/testo 650/testo 900
I/O
Main menu
Probe
Calibration
Calibration
Connect probe to
calibration set
Manual calibration via middle
button "Cal" or automatically
after a certain time (cancel via
"ESC"). Jump back to measure-
ment menu by pressing the
"ESC" button 3x.
These pipes contain a small amount of caustic soda. Take care
when handling the pipe and observe the information on the
label or packaging. In particular, mechanical loads should be
avoided when connecting the pipes. The two plugs on the pipe
connection should be removed before connecting the pipe. These
plugs should be attached to the connections immediately after use.
Do not throw the pipe into your normal domestic rubbish. It should
be disposed of using special disposal services.
• Connect the CO
and the reducing adapter.
• Connect the other side of the CO
output via the connection pipe and reducing adapter
Remove
(green connection).
• Ensure that pump movements are slow because the absorption
material needs a certain time to absorb (gas speed should not be
too high).
• Rinse the probe for a few minutes with CO
testo 454
<OK>
<OK>
<OK>
<OK>
* (do not calibrate with "
This is a 1 point calibration at zero point. CO
supplied to the CO
Measuring range
Accuracy
t 90 (without pump operation) :
Perm. operating temperature : +0 to +40 °C
Perm. storage temperature
Warm up time after switch on : Approx. 1 minute
Permissible operating pressure : 800 to 1100 mbar (abs. pressure)
Factory calibration
Permissible one point
calibration temperature
Stability
absorber pipe to the probe via connection pipes
2
Example:
Selection:
Main menu/Probe
Selection:
Calibration
Selection:
K1:CO
ACTUAL 0.065
SETTING 0
Wait until the actual value is stable
(near zero).
Question:
Do you really want to calibrate?
Note:
Calibration
K1:CO
Return:
3 x <Arrow>
sensor.
2
3
CO
Zero point calibration
absorber pipe to the pump ball
2
free gas.
2
<Enter>
<Enter>
<Enter>
2
<Enter>
<Enter> *
")
2
free gas must be
2
Technical data
: 0 to 1 vol% CO
2
= 0 to 10,000 ppm CO
: ± 5% of meas. value or 50 ppm
60 seconds
: -20 to +50 °C
: scaled to 1013 hPa
(M.S.L.) and 25 ° C
: 15 to 30 ° C
: ± 500 ppm / year
P
ROBE
2
2