Herunterladen Inhalt Inhalt Diese Seite drucken

Storage And Transport; Mounting; General Instructions - VEGA VEGASWING S 52 Betriebsanleitung

Vorschau ausblenden Andere Handbücher für VEGASWING S 52:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Mounting

If a malfunction is detected or in case of power
failure, the electronics takes on a defined switching
condition, i.e. the output transistor blocks or the
contactless switch is open (safe condition).
Functional principle
The tuning fork is piezoelectrically energised and
vibrates at its mechanical resonance frequency of
approx. 1200 Hz. The piezos are fixed mechan-
ically and are hence not subject to temperature
shock limitations. The frequency changes when the
tuning fork is covered by the medium. This change
is detected by the integrated oscillator and
converted into a switching command.

2.3 Storage and transport

Packaging
Your instrument was protected by packaging
during transport. Its capacity to handle normal
loads during transport is assured by a test
according to DIN EN 24180.
The packaging consists of cardboard. This material
is environment-friendly and recyclable. Dispose of
the packaging material via specialised recycling
companies.
3 Mounting

3.1 General instructions

Switching point
In general, VEGASWING S 52 can be installed in
any position. The instrument simply has be
mounted in such a way that the tuning fork is at the
height of the desired switching point.
14
2
1
3
Fig. 11: Vertical mounting
1
Switching point approx. 13 mm in water
2
Switching point with lower density
3
Switching point with higher density
1
2
Fig. 12: Horizontal mounting
1
Switching point
2
Switching point (recommended mounting position, particularly
for adhesive products)
Moisture
Fig. 13: Measures against moisture penetration
VEGASWING S 52

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis