Specifications

Power:
Environmental:
Range:
Beam Angle:
Memory:
Programming:
Temperature Compensation:
Display:
Output
Page 10
Downloaded from
www.Manualslib.com
• 18 to 30 V DC, 0.2 A max
• location:
indoor / outdoor
• altitude:
2000 m max.
• ambient
continuous:- 40 to 60 °C (-40 to 140 °F)
temperature
• relative humidity:
suitable for outdoor (Type 4X / NEMA 4X / IP65 enclosure)
• installation category:II
• pollution degree:
4
• process pressure vented to atmosphere
• 0.25 to 5 m ( 0.8 to 16.4 ft.), liquids only (standard 24 V model, black label)
• 0.25 to 8 m ( 0.8 to 26.2 ft.) (Extended Range model, green label)
• 10° at -3 dB boundary
• non-volatile EEPROM, no battery required
• 2 tactile keys
• built-in to compensate over the operating range.
• liquid crystal
• three 9 mm (0.35") digits for reading of distance between sensor face and material
• multi-segment graphic for operation status
• mA:
range:
span:
accuracy: 0.25% of full scale
resolution: 3 mm (0.125")
loading:
cable:
• Relay:
1 normally closed contact rated at 5 A at 250 V AC non-inductive or 24 V DC
fault on power, application or device failure
The Probe – INSTRUCTION MANUAL
manuals search engine
-20 °C (-5 °F) if metal mounting
4 – 20 mA
proportional or inversely proportional
750 ohms max at 24 V DC supply
Belden 8760, shielded, twisted pair, 28 AWG (0.75 mm
equivalent
2
) or
7ML19985GD61