Calibration
The calibration of the mA output may be done such
that its span will be either proportional or inversely
proportional to the material level.
Calibration: Reference Method
4 mA calibration
20 mA calibration
Operation Status
The graphic portion of the display gives the user a visual
indication of The Probe's operating status. Viewing the
graphic can assist the user in properly locating and
installing The Probe to achieve optimum performance.
Page 4
Downloaded from
www.Manualslib.com
The 4 and 20 mA levels may be calibrated in
Note:
any order.
proportional span
high level = 20mA
low level = 4 mA
• Adjust the material level (or target) to the desired distance from the sensor face.
• Press the "4" or "20" key (as appropriate) to view the stored distance associated with that mA
output value.
• Press the key a second time to set the new distance reference.
• After viewing or calibrating, Probe operation automatically reverts to the Run mode (6 sec). The
calibration value is referenced from the face of The Probe sensor, in the units displayed.
Press "4"
Press "4" again
Press "20"
Press "20" again
Calibration bypasses the measurement response rate.
Note:
The logo will change from full to partial to indicate operation status. After the `Waiting' period, the `?'
icon will appear for an `LOE / FAULT' indication. When a valid echo is again received, a `Good'
indication will resume. Refer to Troubleshooting on page 9 .
The Probe – INSTRUCTION MANUAL
manuals search engine
inversely proportional span
high level = 4 mA
low level = 20mA
4 mA calibration
calibration invalid if
new 4mA calibration
retry
20 mA calibration
calibration invalid if
new 20 mA calibration
retry
GOOD
WAITING
LOE / FAULT
7ML19985GD61