4. Calibration

4.2 Two point calibration (input scaling)

Using the "Start calibration..." button, a new dialog box is displayed which conducts the
calibration process.
GB
Step 1
Initially, the start of the measuring range must be gener-
ated.
Enter the actual pressure present (reference value) in the
input field and confirm using the "Confirm" button.
Step 2
Next the end of the measuring range must be generated.
Enter the actual pressure present (reference value) in the
input field and confirm using the "Confirm" button.
Step 3
By actuating the button „Yes" the parameters are stored.
Pressing „Cancel" will discard all changes.
24
tecsis instruction manual CommunicationSoftware CANopen
®