4. Calibration

4. Calibration
There are two possibilities available for calibration.
Zero adjustment
Two point calibration

4.1 Zero adjustment (Input offset)

The "Read" button will give information about the currently-programmed offset.
For zero adjustment there are two possibilities:
AutoZero: The current pressure will be set as the start of the measuring range.
In order for the AutoZero function to be used with measuring ranges with
negative pressure ranges (e.g. -1 ... 6 bar), the start of the measuring
range must be reached, e.g. -1 bar (not 0 bar)!
Writing the zero adjustment: In the input field, the manually-determined offset for the
pressure sensor can be given and adjusted.
Independent of the unit set on the instrument, a user-defined pressure unit can be set
via the drop-down list.