4.2.2 Vertical Calibration
Vertical calibra-
tion, step by step
Saving the cali-
bration
Checking and Adjusting
1. Turn the laser off before switching to calibration mode. Simultaneously
press On/Off and Auto/Man.
2. After a few seconds, release On/Off key.
3. The X LED indicator (19) will blink, then the Y LED (21). Release the
Auto/Man key.
4. The Z LED (23) will blink rapidly; indicating laser is ready to be calibrated
in Z-axis.
5. Move the beam until it's vertical and parallel to the plumb line using (17)
or (16) on the laser or (2) or (3) on the ZDR300. Move the laser slightly so
that the beam is over the plumb line for the final check.
The laser is now calibrated in Z-axis. Press (14) on laser or (10) on detector
to save the calibration data. If you don't wish to save the calibration, press the
On/Off Key on the laser.