10. Trouble shooting
Please verify in advance if pressure is being applied (valves/ ball valve etc. open) and if the right
voltage supply and the right type of wiring (2-wire/ 3-wire) has been chosen?
Failure
No output signal
No/False output signal
Output signal unchanged after change
in pressure
Signal span dropping off/too small
Signal span erratic / incorrect
Abnormal zero point signal
In case of unjustified reclamation we charge the reclamation handling expenses.
Betriebsanleitung/Operating instructions/Mode d'emploi/Instrucciones de servicio
Downloaded from
www.Manualslib.com

Possible cause

Cable break
No/incorrect voltage supply or current
spike
Incorrectly wired (e.g. Connected as
2-wire instead of 3-wire system)
Mechanical overload through
overpressure
Mechanical overload through
overpressure
Diaphragm is damaged, e.g. through
impact, abrasive/agressive media;
corrosion of diaphragm/pressure
connector; transmission fluid missing.
Seal/Sealing face damaged/conta-
minated, seal mounted incorrectly,
threads crossed
Electromagnetic interference source
in the vicinity, e.g. inverter drive
Working temperature too high/too low
Instrument not grounded
Violent fluctuations in the process
media pressure
Working temperature too high/too low
Abnormal mounting position
Overload limits exceeded
manuals search engine
Procedure
Check connections and cable
Adjust the voltage supply to correspond
with the Operating Instructions
Follow pin assignment (see Instrument
Label / Operating Instructions)
Replace instrument; if failure reoccurs,
consult the manufacturer
Replace instrument; if failure reoccurs,
consult the manufacturer
Contact the manufacturer and replace
the instrument
Clean the seal/sealing face, possibly
replace the seal.
Shield the device; shield the cables;
remove the interference source
Ensure permissible temperatures as per
the Operating Instructions
Ground instrument
Damping; consult with manufacturer
Ensure permissible temperatures as per
the Operating Instructions
Correct the zero point through the poten-
tiometer, control panel or software
Ensure permissible overload limits are
observed (see Operating Instructions)