Herunterladen Inhalt Inhalt Diese Seite drucken

Siemens SITRANS Probe LU Kurzanleitung Seite 14

Vorschau ausblenden Andere Handbücher für SITRANS Probe LU:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 31
Quick Setup: steps 1 to 10
Note:
1. Select language (P010: Language)
Values
2. Set P001: Operation (measurement mode)
Notes:
• Setting P001 resets Span (P007), unless Span has previously been set to a different value.
• Changing P001 may reset Output Function (P201): this applies to HART only.
Values
transducer
face
P006
Empty
3. Set Measurement Response (P003)
Values
Set P003 to a measurement response speed just faster than the maximum filling or emptying
rate (whichever is greater).
Page EN-12
Downloaded from
www.Manualslib.com
Default values (factory settings) are marked by an asterisk (*) in the tables.
Numeric/None
0 *
English
1
German
2
French
3
Spanish
4
Level returns distance to material level referenced from Empty (process empty level). The
1 *
reading is returned in volumetric units if parameters 050 to 055 are set to enable this.
2
Space returns distance to material level referenced from Span (process full level).
3
Distance returns distance to material level from reference point (transducer face).
mA Output with Level, Space, and Distance operation
Level
(P001 = 1)
(20 mA)
Span
100%
P007
Level
0%
(4 mA)
1
*
slow
0.1m/minute
2
medium
1m/minute
3
fast
10m/minute
SITRANS Probe LU (HART) – QUICK START MANUAL
manuals search engine
ENGLISH
DEUTSCH
P000
LOCK
VERRIEGELG
P001
OPERATION
BETRIEB
P003
MEAS RESP
REAKTIONSZ
P004
ANTENNA
ANTENNE
P005
UNITS
EINHEIT
P006
EMPTY
MESSBER.
P007
SPAN
MESSSPANNE
P010
LANGUAGE
SPRACHE
Space
(P001 = 2)
(4 mA)
Span
0%
Space
Empty
100%
(20 mA)
FRANÇAIS
ESPAÑOL
VERROUIL
BLOQUEO
FONCTIONMT
FUNCIONAM.
TEMPS REP .
TIEMPO R.
ANTENNE
ANTENA
UNITES
UNIDADES
VIDE
VACIO
PLAGE
RANGO
LANGUE
IDIOMA
Distance
(P001 = 3)
0%
(4 mA)
Span
Distance
Empty
100%
(20 mA)
A5E32168031

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis