Programming SITRANS LR260
A Quick Start Wizard provides an easy step-by-step guide to help you configure the device for
a simple application.
• 7-step
• 4-step
Settings can be modified locally via the Local User Interface (see
the handheld programmer
Interface (LUI) consists of an LCD display and a handheld programmer.
Activating SITRANS LR260
Power up the device. SITRANS LR260 automatically starts up in Measurement mode. Press
Mode
Page EN-8
Downloaded from
www.Manualslib.com
4.
The equipment is certified for use in an ambient temperature range of -40 C to 80 C.
5.
The equipment has not been assessed as a safety related device (as referred to by
Directive 94/9/EC Annex II, clause 1.5).
6.
Installation and inspection of this equipment shall be carried out by suitably trained
and authorized personnel in accordance with the applicable code of practice.
7.
The equipment shall be installed such that the supply cable is protected from
mechanical damage. The cable shall not be subjected to tension or torque. The
equipment manufacturer is not responsible for providing the supply cable.
8.
Repair of this equipment shall be carried out by suitably trained and authorized
personnel in accordance with the applicable code of practice.
9.
The equipment shall be supplied with an input voltage not exceeding 30 V DC. The
source shall provide electrical isolation between its input and output, meeting the
applicable safety requirements of IEC 61010-1.
SPECIAL CONDITIONS FOR SAFE USE
The 'X' suffix to the certificate number relates to the following special condition(s) for
safe use:
Unused cable entries shall be fitted with blanking elements that can only be
removed with the aid of a tool.
Any glands, conduit entry devices or blanking elements fitted to the equipment
shall be suitable for use in the presence of combustible dusts and certified as
such by a notified body; the installation of these devices shall not compromise
the IP6X rating of the equipment enclosure.
For applications that use the purge facility, the user shall provide a means to
ensure that combustible dust from the hazardous area cannot enter the purge
supply in such a way as to compromise the area classification.
Quick Start Wizard via the handheld programmer
Quick Start Wizard via SIMATIC PDM
on page 10) or remotely via SIMATIC PDM. The Local User
to toggle between Measurement and Program Mode.
SITRANS LR260 (PROFIBUS PA) – QUICK START MANUAL
manuals search engine
on page 13.
on page 16.
Accessing parameters via