Calibration With Bam- And Working-Standard - ColorLite sph860 Bedienungsanleitung

Spektrophotometer
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Device overview
The sensitivity setting should be adapted to the required
precision
[see 6.1.8 Calibration
ColorLite GmbH | sph860/sph900
Downloaded from
www.Manualslib.com
The device will display a message
when the calibration was success-
ful. Return to the main menu by
pressing the "ENTER"
If not successful, an error message
"Calibration is invalid" is dis-
played.
Also the sph860/sph900 has a cal-
ibration warning function. This
informs the user that a recalibra-
tion is needed. The sensitivity of
this function can be adjusted in
"Settings" – "System" –
"Calibration warning".
warning].
manuals search engine

5.1.2 Calibration with BAM- and Working-Standard

If the option "Working-Standard" is activated the cali-
bration menu of the sph860/sph900 will show a second
key.
menu option - "Working-Standard" - as well as the
"BAM-Standard".
It is important that the two white standards are
measured in the correct order.
Measure the BAM standard and then directly the working
standard in one procedure.
Select "BAM-standard" by pressing the "ENTER"
key. Follow the displayed instructions:
Connections
First calibrate with the BAM-
Standard.
Second you will be prompted to
calibrate with the Working-
Standard.