6.1.7 Using a working standard

To protect the sensitive BAM white standard, ColorLite
supplies robust working-standards as an option. These
working standards have a glossy surface and can be
cleaned using an alcohol cleaner, such as isopropyl.
In "Settings" - "System" switch on the option "Use
working standard". This mode can cause confusion
when the wrong standard is used to calibrate and is why a
warning message is displayed.
When activated the calibration menu of the
sph860/sph900 will show a second menu option "Work-
ing standard" as well as the "BAM Standard".
Instructions for using this function see also in chapter
5.1.2
Calibration with BAM- and
ColorLite GmbH | sph860/sph900
Downloaded from
www.Manualslib.com
Fig. - Different
working standards
Working-Standard.
manuals search engine

6.1.8 Calibration warning

The
sph900
has a calibration warning function. This
informs the user that a recalibration is needed.
The sensitivity of this function can
be adjusted in "Settings" – "Sys-
tem" – "Calibration warning".
The sensitivity can be selected as
an index of 0 (off) to 9 (most sen-
sitive setting). Select the desired
level with the arrow keys
confirm with "Enter".

Time interval

A time interval warning since the last calibration can also
be set in intervals up to 48h or switched off.
The sensitivity setting should be adapted to the re-
quired precision – The advised setting for normal
use is 3 and 24h.