Herunterladen Inhalt Inhalt Diese Seite drucken

wtw LF 320/SET Bedienungsanleitungen Seite 36

Standard-leitfähigkeitsmeßzelle/mikroprozessor konduktometer
Vorschau ausblenden Andere Handbücher für LF 320/SET:
Inhaltsverzeichnis

Werbung

Salinity
TDS
Temperature
Data storage
Display
Keyboard
Inputs
Environmental conditions
Ambient temperature
Relative humidity
EMC
Interference transmission
Disturbance
36
Downloaded from
www.Manualslib.com
Range:
Resolution:
Accuracy:
Range:
Resolution:
Accuracy:
Range:
Resolution:
Accuracy:
manuals search engine
Technical data
0.0 ... 70.0
0.1
after adjustment (via cell constant) to 35.0:
Ÿ ± 0.1 in the range 0.0 ... 42.0 at test sample
temperature 5 ... 25°C
Ÿ ± 0.2 in the range of 0.0 ... 42.0 at test sample
temperature 0 ... 5°C resp. 25 ... 30°C
Ÿ better than 3 % of meas. value in the range
42.0 ... 70.0
0 ... 1999 mg/l
1 mg/l
± 2 % ± 1 digit at test sample temperature
15°C ... 35°C.
Factor adjustable between 0.40 ... 1.00.
-5 ... 99.9°C
0.1 K
0.1 K ± 1 digit at operating temperature
-10°C ... +55°C
Ring memory
Number of values: 50 pairs
Output:
Display
Data protection: any long
WTW-specific LCD-display
Height:
35 mm
Width:
60 mm
Field with 9 keys
8-pin WTW-plug
Operation:
- 10°C ... + 55°C
Storage:
- 25°C ... + 65°C
Yearly mean:
< 80 %
30 days/year:
< 100 %
Remaining days: < 90 %
Technical fundamental standard EN 50081-1
FCC class A
Technical fundamental standard EN 50082-2
Namur recommendation (standard requirements)
BA52207/03.95/Pap/LF320-SET-1

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Lf 320Tetracon 325

Inhaltsverzeichnis