Ex-classification:
II 1/2G Ex op is / op pr IIA 95 °C Ga/Gb
II 1/2G Ex op is / op pr IIB 38 °C Ga/Gb
Examination certificate number: SEV 12 ATEX 0169 X
Notified body: Electrosuisse SEV, (No.: 1258)
Technical Data
Technical
InPro 8200/S(H) InPro 8100/S
data
Meas.
dual fiber
principle
Shaft
DIN 1.4435 or
material
1.4404 (316L)
or Hastelloy
C276
(available only
with Epoxy
sealed window)
Shaft
12 mm
diameter
Shaft
120, 205,
length
297 or 407 mm
Window
Sapphire
material
O-ring
Viton
sealing
Process
Pg 13.5
adaption
Plug
2  SMA
Meas.
5 FTU...4000 FTU
range
0...30 g/l
Pressure
0...6 bar
range
(0...87 psi) with
Kalrez
sealed window
0...16 bar
(0...232 psi)
with Epoxy bon-
ded window
Tempera-
–10...130 °C
ture range
(14...266 °F)
with Kalrez-FDA
sealed window
– 30...130 °C
(– 22...266 °F)
with Epoxy bon-
ded window
Steam-
130 °C (only
steriliz-
with Kalrez-FDA
able
sealed window)
Subject to technical changes.
4
Sensor installation
The turbidity sensors are to be installed at a distance of at
least 10 cm (4") from the walls of the vessel, agitator or from
the pipe walls in order to prevent disturbances through re-
flections. If it is not feasible in pipes of small diameter then
the sensor can be installed sloped, at an angle of 15...45°
or, alternatively, in a knee bend. The pipe must always be
completely full of liquid.
5
Operation
If the turbidity measuring system is calibrated off-line, it is
recommended to use the METTLER TOLEDO calibration tool,
CaliCap
shaft. If a CaliCap is not used, a minimum distance of 10 cm
(4") to the vessel walls has to be observed as well. During
calibration, the sample should be stirred in order to prevent
the sedimentation of solid particles. In case of any failure or
InPro 8000 Series
30 313 267 B
®
-FDA
®
-FDA
®
, (ordering no. 52 800 210) for fitting on the sensor
single fiber
1.4435 or
1.4404 (316L)
12 mm
120, 205,
297 or 407 mm
without window
Viton-FDA
Pg 13.5
2  SMA
10 FTU...4000 FTU
0...250 g/l
0...6 bar
(0...87 psi)
– 30...130 °C
(– 22...266 °F)
130 °C
© 07 / 2019 Mettler-Toledo GmbH
InPro 8000 Series
InPro 8050
single fiber
PSU
(Polysulfone)
12 mm
120 mm
Sapphire spigot
Viton-FDA
Pg 13.5
2  SMA
10 FTU...4000 FTU
0...250 g/l
0...2 bar
(0...29 psi)
0...60 °C
(32...140 °F)
non-sterilizable