Directions for use
The handle and charging accessory is intended for use in the electromagnetic environment specified below. The
customer or the user of the handle and charging accessory should assure that it is used in such an environment.
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
Power frequency (50/
60 Hz) magnetic field
IEC 61000-4-8
Note: U
is the a.c. mains voltage prior to application of the test level.
T

Electromagnetic immunity

IEC 60601 test level
Compliance level
±8 kV contact
±8 kV
±15 kV air
±15 kV
±2 kV for power supply
±2 kV
lines
±1 kV for input/output
±1 kV
lines
±0.5 kV, ±1 kV
±1 kV
Line- to -line
±0.5 kV, ±1 kV, ±2 kV
±2 kV
Line-to-ground
0 % U
; 0.5 cycle
0 % U
; 0.5 cycle
T
T
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
0 % U
; 1 cycle
0 % U
; 1 cycle
T
T
70 % U
; 25/30 cycles
70 % U
T
T
Single phase: at 0°
0 % U
; 250/300 cycle 0 %U
; 250/300 cycle
T
T
30 A/m
30 A/m
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment. If the user
of the handle and charging accessory requires
continued operation during power mains
interruptions, it is recommended that the
handle and charging accessory be powered
from an uninterruptible power supply or a
battery.
; 25/30 cycles
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.