SPECIFICATIONS (CONTINUED)
English
Recommended separation distances between portable and
mobile RF communications equipment and the SMRT Charger
The SMRT Charger is intended for use in the electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the SMRT Charger can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and
the SMRT Charger as recommended below, according to the maximum output power of the communications
Rated maximum
output power of
transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
Return To Table of Contents
1-42
EMC Information
equipment.
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d=1.17√P
0.12
0.37
1.2
3.7
12
6500-109-105 REV A
m
80 MHz to 800 MHz
d=1.17√P
0.12
0.37
1.2
3.7
12
800 MHz to 2.5 GHz
d=2.33√P
0.23
0.74
2.3
7.4
23
www.stryker.com