INSTALLING THE OPTIONAL MOUNTING BRACKET (CONTINUED)
2.
Position the mounting bracket, by making sure that the:
• Spring tab is located at the rear of the SMRT Charger.
Note: When mounting the bracket to a vertical surface, the spring tab
should be horizontal with the spring tab pointing to the left as shown in
Figure 2 on
• Power cord easily plugs into the rear of the SMRT Charger.
• SMRT Charger slides from front to back to engage the bracket after
mounting.
• Bracket is installed correctly for the ambulance or station location:
Location
Ambulance • Secure the bracket to a horizontal structural support
Station
WARNING
Ensure that the mounting bracket is securely attached to the surface.
www.stryker.com
Charger Installation
page
1-16.
Considerations
surface or shelf using size #10, grade 5 screws minimum
(not supplied).
• Make sure that the selected mounting surface is strong
enough to securely hold the SMRT Charger and SMRT
Pak during transport.
• Locate the bracket in an enclosed cabinet and out of
patient reach during transport.
• Allow for easy SMRT Pak installation and removal.
• Locate the power supply within reach of the power cord.
• Secure the bracket to a horizontal or vertical structural
support surface using size #10, grade 5 screws minimum
(not supplied).
• For vertical installation, position the mounting bracket,
so the spring tab is directly below the mounting screws
to ensure that the SMRT Pak remains supported by the
SMRT Charger even if the release button is accidently
pressed.
• Allow for easy SMRT Pak installation and removal.
• Locate the power supply within reach of the power cord.