Chapter 3. Operating precautions,
checks and warnings
Warnings:
Electrical safety testing should be performed by a
biomedical engineer or other qualified person.
Risk of electrical shock: do not remove the cover.
Refer servicing to qualified or authorized FMS Repair
Service Center.
Do not connect the device to a power source that
is not properly earthed (grounded).
Disconnect the device from the main power source
when cleaning, servicing, or inspecting.
Inspect equipment and cables periodically for
wear. Return to FMS Repair Service Center if
damage is noted.
Avoid fluid contact with the console, and its
electrical connectors.
D o not use flammable agents when cleaning and
disinfecting the console.
To avoid risk of fire, replace fuses with same type
and rating.

.1. Sterile Packaging:

Carefully examine the shipping package and sterile
wrap. If the package is damaged, the sterile seal
is broken, or if the Expiration Date is passed,
do not use.

.2. Storage and Handling

Storage temperature: Do not expose the
console to temperatures above 50°C (122°F) or
below –10°C (14°F).
Operating Conditions: Do not operate in
conditions above 50°C (122°F) or below
–10°C (14°F).
Damage: Do not use the fms duo
shaver if dropped or shows signs of damage.
Disposal: Do not incinerate.
Return to FMS for safe disposal.
Factory Settings: Factory settings are based on
experience they should only be used as guidelines.
The surgeon is responsible for settings depending
on the surgical procedure performed.