cleaning and disinfection of the turbine
cleaning and disinfection of the mouthpiece
cleaning of the device
replacing batteries
3.1
Cleaning and disinfection of the turbine
Correct operation of the turbine is guaranteed only if it is "clean" and free of foreign objects
that affect its movement. The presence of dust or foreign bodies (such as hairs, sputum etc.)
could slow or block the moving parts of the turbine and make the result less accurate, or
damage the turbine itself.
After each use, check the cleanliness of the turbine.
To clean the turbine, pull it out of the SMART ONE socket by turning it counterclockwise and
simply pulling it out. To make it easier to pull out, push the base of the turbine gently with a
finger.
Soak the turbine in a cold detergent solution and shake it to remove any impurities inside it;
leave it to soak for the length of time suggested by the cleaning solution manufacturer and
shown in the operating instructions.
Rinse the turbine by dipping it in clean cold water.Drain the turbine and shake it vigorously.
Leave it to dry by laying it with the shaft vertical to the dry support surface.
After cleaning, insert the turbine into the socket in the direction indicated by the screen-
printed closed padlock symbol on the SMART ONE. To insert the turbine correctly, push it
down and turn clockwise until it stops, to make sure it is fully inserted into the plastic
container.