Herunterladen Inhalt Inhalt Diese Seite drucken

Omron HBP-1100 Gebrauchsanweisung Seite 7

Vorschau ausblenden Andere Handbücher für HBP-1100:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 86
Caution
• Before using the unit, verify that none of the following apply to the
patient:
- Poor peripheral circulation, noticeably low blood pressure, or low body
temperature (there will be low blood flow to the measurement position)
- The patient uses an artificial heart and lung (there will be no pulse)
- An SpO
sensor and the cuff are attached to the same arm
2
- The patient has an aneurysm
- The patient has frequent arrhythmia
- Body motions such as convulsions, arterial pulsations, or trembling
(cardiac massage in progress, minute continuous vibrations,
rheumatism, etc.)
• Before use, visually inspect the unit to make sure there are no
deformations due to falling, and that there is no dirt or moisture on the
unit.
• When the unit has not been used for an extended period of time, always
verify that it operates normally and safely before use.
• Do not use in a location where the unit may easily fall. In the event that
the unit falls, verify that it operates normally and safely.
Cleaning
Warning
• When cleaning the unit, turn off the power and disconnect the AC
adapter from the unit.
• After cleaning the unit, make sure it is completely dry before connecting
to a power outlet.
• Do not spray, pour, or spill liquids into or onto the unit, accessories,
connectors, buttons, or openings in the housing.
Caution
• Do not use thinner, benzene, or other solvents to clean the unit.
• Do not sterilize by autoclave or gas sterilization (EOG, formaldehyde
gas, high-concentration ozone, etc.).
• If using an antiseptic solution for cleaning, follow the instructions of the
manufacturer.
• Clean the unit regularly.
EN
5

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis