Remove the transportation lock:
On the left lower side of the balance replace the golden-coloured screw with the
silver-coloured one (see sticker).
7.3 Mains connection
Select a country-specific power plug and insert it in the mains adapter.
Check, whether the voltage acceptance on the scales is set correctly.
Do not connect the scales to the power mains unless the information on
the scales (sticker) matches the local mains voltage.
Only use KERN original mains adapter. Using other makes requires
consent by KERN.
Important:
Before starting your weighing balance, check the mains cable for
damage.
Ensure that the power unit does not come into contact with liquids.
Ensure access to mains plug at all times.
In order to obtain exact results with the electronic balances, your balance
must have reached the operating temperature (see warming up time chap. 1).
During this warming up time the balance must be connected to the power
supply (mains, accumulator or battery).
The accuracy of the balance depends on the local acceleration of gravity.
Strictly observe hints in chapter Adjustment.
7.4 Connect and switch-on power supply
Connect scale to power supply.
The balance will carry out a self-test Then the balance is set into stand-by mode.
The weighing balance will remain switched on as long as it is connected to the
power supply. Operate the ON/OFF-key to turn on/off the display.
32
.
ALJ/ALS/PLJ/PLS-BA-e-1944