Press On/Off to advance to the Setup Menu, or press Tare to return to Span
calibration.
Note:
To exit a calibration at any time, press and hold the On/Off key until the
scale shuts off.

4.2 Setup Menu

Enter this menu to set balance parameters.
Light:
Unit:
Auto shut-off:
End
Note: Units vary for different regions or models.
L.I.G.H.t
Light [
]
This sets the operation of the display's backlight.
ON = always on
OFF = always off
AUTO = Turns on when a button is pressed or the displayed weight changes.
Alternatively, turns off the backlight when the displayed weight does not change for
more than 10 seconds.
U.N.I.t
Unit [
]
The Valor 1000 allows one weighing unit to be used at a time. Units can be
changed in this menu setting and will appear when you exit the menu.
Press On/Off to enter the Unit submenu. You will see "kg" displayed for kilograms.
Press On/Off to accept kilograms as the weighing unit and exit the Unit submenu,
or press Tare to move to the next unit.
Auto Shut-off Menu [
Set the automatic shut off time for the scale.
A00 = powers off after 5 minutes of no activity with battery powered only; disabled
with AC adapter powered
A01 = powers off after 1 minute of no activity
A05 = powers off after 5 minutes of no activity
A10 = powers off after 10 minutes of no activity
Off = disabled
Off, On, Auto
kg, g, lb, oz, lb:oz
A00, A01, A05, A10, Off
Exit menu