INSTALLATION – OPERATION – MAINTENANCE
within about 6 – 10 hours). Suction tips holder is equipped with vacuum regulation of suction
unit depends on dentist needs.
The state of filling of the separation vessel is evaluated by the separation automatics which
interrupts the suction by the aggregate (9) and filling of the separation vessel is indicated
automatically with the indicator lamp (4) going on. At this point it is necessary to put the
tubings with suction mouthpieces into the holder and switch the power-supply switch off.
Afterwards open the door by pulling the grips situated on the side walls, uncouple rubber
fixing lugs from the separator cover (10) and pull the separator's vessel out. Hang the
separator's cover into the holder on the left side (8).
Empty the separation vessel content into the sewer, rinse the vessel with water, connect with
the separator's cover in a reversal order. Set the separator into a bowl (12) in the aspirator
box (the bowl is detachable). Check the connection of the cover with the vessel, seating of
the separation vessel in the bowl and close the door.
The tips are not delivered in a sterile state!
Replace used suction mouthpieces on tubings after each patient!
The used tips that are designed for multiple use are necessary to be sterilized
after each patient !
The cannulas are to be exchanged after maximum of 400 cycles of
sterilisation or after one year based on whichever occurs first.
It´s forbidden to cover all blow-holes placed on upper sides of unit.
Aspirator contains no spare power supply!
Before each switch on running unit it´s neccessary to check if door of box is
closed.

9. MAINTENANCE INTERVALS - USER / TECHNICIAN

Maintenance to be done
Inlet Sieve Cleaning
Disinfection of Tubings and
Separation Vessel
Replacement of output pre-filter
Replacement of output filter
Exchange of suction cannula
Regular inspection
* Note: the suction end-pieces are exchanged at one of these events, whichever occurs first.
NP-DO M-19_11-2010.doc
Chapter
Time interval
10.1
per day
10.2
per day
10.3
every 3 months
10.4
per annum
Maximum of 400
cycles of sterilisation
or once a year *
1x in 2 years
- 21 -
ASPINA DO M
To be performed by
user
user
user
user
user
service technician