In general, when used in those concentrations recommended by the manufacturer, either phenolic type or quaternary
(excluding Virex
because staining may result.
Suggested cleaners for the 6250, 6251, and 6252 surfaces:
• Quaternary Cleaners (active ingredient - ammonium chloride)
• Phenolic Cleaners (active ingredient - o-phenylphenol)
• Chlorinated Bleach Solution (5.25% - less than 1 part bleach to 100 parts water)
Avoid over saturation and ensure the product does not stay wet longer than the chemical manufacturer's guidelines
for proper disinfecting.
WARNING
• SOME CLEANING PRODUCTS ARE CORROSIVE IN NATURE AND MAY CAUSE DAMAGE TO THE PRODUCT IF
USED IMPROPERLY. If the products described above are used to clean Stryker patient care equipment, measures
must be taken to insure the chairs are wiped with a cloth soaked in clean water and thoroughly dried following
cleaning.
• Failure to properly rinse and dry the chairs will leave a corrosive residue on the surface of the chairs, possibly
causing premature corrosion of critical components.
Note: Failure to follow the above directions when using these types of cleaners may void this product's warranty.
Return To Table of Contents
1-30
®
TB) type disinfectants can be used. Iodophor type disinfectants are not recommended for use
Cleaning
6250-001-161 REV D
www.stryker.com