4. Gravitation constant at place of installation
5. Gravitation constant applied during verification
Press
repeatedly to exit menu.
15.4.1 Configuration example single range scale
Invoke menu item<mult> (see chap. 15.4) and press
confirm. The currently set balance type is displayed.
Press
to select desired type
SinGLE = single range scale
Press
, the screen used to enter readability/verification
value will be shown.
Press
, the current setting will be displayed.
Select desired setting with
Press
to select next menu item for entering capacity.
Press
, the current setting will be displayed.
Press
to select the desired setting.
Confirm by
, „End" will be shown.
To import the configuration data, press
The entering of configuration data requires subsequent
adjustment or linearization.
For how to carry out adjustment see chap. 7.7 or
linearization see chap. 7.8
KXS_KXG-TM-BA_IA-e-1932
and acknowledge by
again.
to
.
65