Herunterladen Inhalt Inhalt Diese Seite drucken

Humidity Measuring Transducer (600X And 602X Series); Humidity Measuring Transducer (60Xx Series) - testo 105679 Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Humidity measuring transducer (600x and
602x series)
Mount the screw connection on the control container.
Connect the control container to the humidity probe
and tighten the rubber bushing.
Mount the probe directly on the humidity adjustment
container.
Remove the cover on the measuring transducer hous-
ing.
You will find 3 switches under the large transducer
cover.
J1
You can adjust the 11.3 %RH point by simultaneously
pressing the S1+S3 buttons, S2 with S3 sets the
75.3 %RH point.
16
Digital adjustment
h h y y g g r r o o t t e e s s t t 600
h h y y g g r r o o t t e e s s t t 602
S1
S2
S3
J2 J3
J4
Adjustment
Digital adjustment
Humidity transducer (6337.9741)
Remove the screwed sleeve on the handle.
Digital adjustment as for humidity probe. Description
on page 15.
Humidity measuring transducer, old version
The measuring transducers from the hygrotest 55
series (item no.0555.60xx) are equipped with a micro-
processor which facilitates adjustment. Period of sta-
bilization: 24 hours.
Remove the cover from the measuring transducer
housing. 2 pushbuttons can be seen. The adjustment
is carried out by pressing the appropriate button.
1.
Press S1.
The current output in the version 0 to 20 mA is
set at 2.4 mA (in the version 4 to 20mA at
5.92 mA). This corresponds to a display of 12%
± 2 digits.
11.3%
2.
Press S2.
75.3%
The current output in the version 0 to 20 mA is
set at 15.2 mA (in the version 4 to 20mA at
16.16 mA). This corresponds to a display of
76% ± 2 digits.
Adjustment
Adjustment
17

Werbung

Kapitel

Inhaltsverzeichnis
loading