Herunterladen Inhalt Inhalt Diese Seite drucken

Baxtran SCU Bedienungsanleitung Seite 36

Kontrollwaage
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

LINEAL CALIBRATION
DISPLAY
LF3
W0
W1
W2
W3
SPEED OF THE AD CONVERTER
SPEED 1 standard speed 15Hz.
SPEED 2 high speed 30Hz.
SPEED 3 low speed 7.5Hz
*This function stays blocked when UF-5 is in mode HOLD 1.
*The value of factory is 1
DISPLAY
LF4
36
After introducing the parameters LF2, the indicator will show the last configuration saved. All the
steps to follow must be completed, if they are not done the indicator will continue with the previous
configuration.
Proceed with the calibration of weight after LF2.
LF 3
DESCRIPTION AND SEQUENCE OF USE
•Press the Key
to start or the key
tically
•Press the Key
to calibrate the zero.
( press the key
o exit from the calibration and go back to the menu LF3)
• Place 1/3 of the maximum weight on the plate and press
the key
to exit from the calibration and go back to the menu LF3)
•Place 2/3 of the maximum weight on the plate and press
(press the key
to exit from the calibration and go back to the menu LF3)
•Place the maximum weight on the plate and press the key
the key
to exit from the calibration and go back to the menu LF3)
LF 4
DESCRIPTION AND SEQUENCE OF USE
•Press the key
to start or the key
again automatically.
USER'S MANUAL SCU
To exit the menu and the scale will start again automa-
to calibrate 2/3 th the capacity. (press
to calibrate 2/3 th the capacity.
to calibrate the full capacity. (press
to exit from the menu and the scale is going to begin
EN

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis