CALIBRATION OF THE WEIGHT
DISPLAY
LF 1
mum capacity and it must also never be exceeded.
CAL Z
•Press the key
the menu LF1)
150.00 kg
•Use
•Place the required weight on the scale as it is indicated in the display.
150.00 kg
150.00 kg
•Once everything is steady, press the key
and go back to the menu LF1).
THE CALIBRATION IS GOING TO FINISH AND THE SCALE WILL GO VACK TO THE WEIGHING MODE AUTOMATICALLY.
CONFIGURATION
LF 2
DISPLAY
FIRST STEP
262144
•DISPLAY OF THE INTERNAL COUNTING
•POSSIBLE VALUES OF THE PARAMETERS
SECOND STEP
A: Metric system
B: American system
C: other unities
1 0 0 0 0 1
D: PCS
A B C D E F
E: double range
F: units of calibration
THIRD STEP
•Use
000000kg
on to the next parameter.
FOURTH STEP
•Use the Keys
dP 0.0
FIFTH STEP
•Use
diV 01, diV 02, diV 05, diV10, diV 20, diV 50
diV 01
•Press
LF 1
DESCRIPTION AND SEQUENCE OF USE
The calibration can be done with any weight, but the weight ca not be inferior to 1/100 of the maxi-
to start the calibration of zero ( press
and
to introduce the weight with which the calibration will be done. ( press the key
to exit the calibration and go back to the menu LF1)
DESCRIPTION AND SEQUENCE OF USE
0:NO
0:NO
0:NO
0:OFF
0:OFF
lb oz cannot be selected as unities of calibration.
The scale won't let us continue up to the next step if there is a mistake during the programming.
and
, to introduce THE MAXIMUM CAPACITY, with the digits. Press
to change the position of the DECIMAL POINT.
d 0.00
d 0.000
to select the DIVISION:
to confirm.
To exit the calibration and go back to
to calibrate it ( press the key ESC to exit the calibration
1: kg
2:T
1: lb
2:lb oz
1: TW kg
2:HK kg
1: ON
1: multi interval
2:multi range
1:use metric
2:use american unities
unities
d 0.0000
USER'S MANUAL SCU
3:g
3:VISS
to move
0.00000
EN
35